Main Unit, Portrait, 100 L/min, NPN. FD-MZ100ATK. Series: Non-Wetted Electrode Electromagnetic Flow Sensors. Series code: FD-M series

Specification

Model NameMain Unit, Portrait, 100 L/min, NPN
Model CodeFD-MZ100ATK
Series TitleNon-Wetted Electrode Electromagnetic Flow Sensors
Series CodeFD-M series
Data Sheet URLhttps://www.keyence.com/download/download/confirmation/?dlSeriesId=WS_SR_fd_m&dlModelId=PM_16Z1ATK&dlSiteId=1&dlLanguageId=en-US&dlLangType=en-US
CAD URLhttps://www.keyence.com/products/process/flow/fd-m/downloads/?mode=cd&modelId=PM_16Z1ATK
Manuals URLhttps://www.keyence.com/products/process/flow/fd-m/downloads/?mode=ma&modelId=PM_16Z1ATK
ModelFD-MZ100ATK*1
TypeVertical
ShapeBuilt-in amplifier
Rated flow range1.4 – 26.0 G/min (5.0 – 100 L/min)*1
Possible display range1.4 – 52.8 G/min (5.0 – 200 L/min)*2
Possible ranges to set0 – 52.8 G/min (0 – 200 L/min)
Connection portNPT1 (25 A)
Detectable fluidsWater or non-corrosive liquid
Conductivity of detection fluids5 µS/cm or higher
Detection fluid temperature0 to +85°C 32 to 185 °F (No freezing)
Pressure lossMax. 1.45 PSI (0.01 MPa)
Display methodDual row display with 4-digit, 7 segment LED, bar display (2 colors), output indicators, flow indicator
Display resolution0.1/1 (G/min, L/min) selectable
Repeatability0.5 s : ±5% of F.S., 1 s : ±3.5% of F.S., 2.5 s : ±2.5% of F.S., 5 s : ±1.6% of F.S., 10 s : ±1% of F.S., 30 s : ±0.8% of F.S., 60 s : ±0.6% of F.S.*3
HysteresisVariable
Response time (chatter prevention feature)0.5 s/1 s/2.5 s/5 s/10 s/30 s/60 s variable
Accumulated flow unit0.1/1/10/100/1000 (G, L) selectable
Integration data storage cycleSave to memory every 10 seconds
Memory backupEEPROM (Data storage length : 10 years or longer, Data read/write frequency : 1 million times or more)
Analog output4-20 mA, max. load resistance 260 Ω Analog output range can be set to any value
Integration reset/bank switch input/flow zero function inputInput time : 20 ms or greater,Select either the accumulation output or the accumulation pulse by setting the mode
Power consumption (Current consumption)Normal : 1700 mW (70 mA), Power save : 1000 mW (40 mA)
Rating24 VDC, Ripple (P-P) 10 % or less, Class 2
Environmental resistance - Enclosure ratingIP65
Environmental resistance - Pressure resistance290 PSI (2.0 MPa)
Environmental resistance - Applied pressure rangeMax. 145 PSI (1.0 MPa)
Environmental resistance - Ambient temperature0 to +50 °C 32 to 122 °F (No freezing)
Environmental resistance - Relative humidity35 to 85 % RH (No condensation)
Environmental resistance - Vibration resistance10 to 55 Hz, Double amplitude 1.5 mm 0.06″, 2 hours in each of the X, Y, and Z directions
Material - Liquid end materialsBore : SCS13, Measurement pipe : PPS, O ring : fluoro-rubber (FKM)
Material - Other materialsPlastic case areas : PPS, Metal case areas : SUS430, Cable : PVC
AccessoryInstruction Manual, Connector cable (2.7 m 106.3″)
WeightApprox. 1,340 g