Fast delivery within 72 Hours

B&R X67AT1322

4 inputs for resistance temperature measurementPT100, PT1000 and other resistance temperature sensorsDirect resistance measurementConfigurable sensor type per channel2-wire and 4-wire measurement…

Request for Quote

Shipping & Delivery

  • Courier delivery

Our courier will deliver to the specified address

5-6 Days

From €20

  • DHL Courier delivery

DHL courier will deliver to the specified address

2-3 Days

From €40

  • Warranty 1 year
  • Free 30-Day returns

Description

4 inputs for resistance temperature measurementPT100, PT1000 and other resistance temperature sensorsDirect resistance measurementConfigurable sensor type per channel2-wire and 4-wire measurement

The module is a temperature module for KTY10, KTY84, Pt100 and Pt1000 resistance temperature sensors. The sensors can be connected with 2-wire or 4-wire connections. The sensor type can be set selectively for each input.

Specification

I/O module4 inputs for KTY10-6, KTY84-130, Pt100 or Pt1000 resistance temperature measurement
B&R ID code0x1488
Status indicatorsI/O function per channel, supply voltage, bus function
Inputs4x M12, A-coded
I/O power supplyM8, 4-pin
Internal I/O1.5 W
CEYes
UKCAYes
CRA (Cyber Resilience Act)In preparation
ATEXZone 2, II 3G Ex nA IIA T5 GcIP67, Ta = 0 – Max. 60°CTÜV 05 ATEX 7201X
ULcULus E115267Industrial control equipment
HazLoccCSAus 244665Process control equipmentfor hazardous locations Class I, Division 2, Groups ABCD, T5
KCYes
Nominal voltage24 VDC
Voltage range18 to 30 VDC
Integrated protective functionReverse polarity protection
InputResistance measurement with constant current supply for 2- or 4-wire connections
Digital converter resolution16-bit
Filter timeConfigurable between 2 and 20 ms
Same sensor types75 ms per channel with 50 Hz filter
When switching sensor type195 ms per channel with 50 Hz filter
Conversion procedureSigma-delta
Output formatINT or UINT for resistance measurement
Sensor typeConfigurable per channel
KTY10-61 LSB = 0.1°C
KTY84-1301 LSB = 0.1°C
Pt1001 LSB = 0.1°C
Pt10001 LSB = 0.1°C
Resistance measurement range0.1 to 4500 Ω / 0.05 to 2250 Ω
Sensor standardEN 60751
Common-mode range±1 VDC
Insulation voltage between channel and bus500 Veff
Linearization methodSoftware
Measuring current250 µA ±1.25%
Reference4530 Ω ±0.1%
Permissible input signalShort-term max. ±30 V
Gain0.01%
Offset0.015%
Max. gain drift0.003 %/°C
Max. offset drift5.25 mΩ/°C
Nonlinearity<0.002%
Crosstalk between channels<-70 dB
G = 11 LSB = Typ. 69.1223 mΩ ±0.1%
G = 21 LSB = Typ. 34.5611 mΩ ±0.1%
Cutoff frequency115 Hz / First-order filter
Slope-20 dB
50 Hz>70 dB
DC>70 dB
Range undershoot0x0000
Range overshoot0xFFFF
Open circuit0xFFFF
General fault0xFFFF
Open inputs0xFFFF
Electrical isolationChannel isolated from busChannel not isolated from channel
AnyYes
0 to 2000 mNo limitation
>2000 mReduction of ambient temperature by 0.5°C per 100 m
Degree of protection per EN 60529IP67
Operation-25 to 60°C
Derating
Storage-40 to 85°C
Transport-40 to 85°C
Width53 mm
Height85 mm
Depth42 mm
Weight200 g
M8Max. 0.4 Nm
M12Max. 0.6 Nm