Fast delivery within 72 Hours

Keyence NR-TH08

High-accuracy temperature/voltage measurement unit. NR-TH08. Series: Multi-input Data Logger. Series code: NR-X series

Request for Quote

Shipping & Delivery

  • Courier delivery

Our courier will deliver to the specified address

5-6 Days

From €20

  • DHL Courier delivery

DHL courier will deliver to the specified address

2-3 Days

From €40

  • Warranty 1 year
  • Free 30-Day returns

Description

High-accuracy temperature/voltage measurement unit. NR-TH08. Series: Multi-input Data Logger. Series code: NR-X series

Specification

Model Name

High-accuracy temperature/voltage measurement unit

Model Code

NR-TH08

Series Title

Multi-input Data Logger

Series Code

NR-X series

Data Sheet URL

https://www.keyence.com/download/download/confirmation/?dlSeriesId=WS_SR57362&dlModelId=PM_117TH08&dlSiteId=1&dlLanguageId=en-US&dlLangType=en-US

CAD URL

https://www.keyence.com/products/daq/data-loggers/nr-x/downloads/?mode=cd&modelId=PM_117TH08

Manuals URL

https://www.keyence.com/products/daq/data-loggers/nr-x/downloads/?mode=ma&modelId=PM_117TH08

Model

NR-TH08 *1

Type

High-accuracy temperature/voltage measurement unit

Input method

Floating unbalanced input, channels insulated from each other, CH inputs insulated from other units (PC),channels not insulated from each other for resistance thermometers

Number of channels

Input: 8 *2

Measurement cycle

Fastest sampling cycle: 100 ms (A/D integration time: 2 ms) to 1 h

A/D converting system

Delta-sigma method

A/D resolution

24 bit

A/D integration time

2 ms, 16.7 ms, 20 ms

Input type

Voltage: ±50 V, ±10 V, ±5 V, ±1 V, ±0.5 V, ±0.1 V; Thermocouple; K, J, E, T, R, S, B, N, C;Resistance thermometer: Pt100, JPt100 3-line type; Specified current: 1 mA

Measurement range - ±50 V

Measurable range: −55.00 to +55.00 VMeasurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±2 digitsDisplay resolution: 10 mV

Measurement range - ±10 V

Measurable range: −11.000 to +11.000 VMeasurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±2 digitsDisplay resolution: 1 mV

Measurement range - ±5 V

Measurable range: −5.500 to +5.500 VMeasurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±2 digitsDisplay resolution: 1 mV

Measurement range - ±1 V

Measurable range: −1.1000 to +1.1000 VMeasurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±2 digitsDisplay resolution: 0.1 mV

Measurement range - ±500 mV

Measurable range: −550.0 to +550.0 mVMeasurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±2 digitsDisplay resolution: 0.1 mV

Measurement range - ±100 mV

Measurable range: −110.00 to +110.00 mVMeasurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±2 digitsDisplay resolution: 0.01 mV

Measurement range - K

Measurable range: −100 to +1372°C -148°F to +2501.6°F, −200 to −100°C -328°F to -148°FMeasurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±0.6°C ±1.1°F, ±0.05% of rdg ±0.9°C ±1.6°FDisplay resolution: 0.05°C 0.09°F*3

Measurement range - J

Measurable range: −100 to +1200°C -148°F to +2192°F, −200 to −100°C -328°F to -148°FMeasurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±0.6°C ±1.1°F, ±0.05% of rdg ±0.8°C ±1.4°FDisplay resolution: 0.05°C 0.09°F*3

Measurement range - E

Measurable range: −100 to +1000°C -148°F to +1832°F, −200 to −100°C -328°F to -148°FMeasurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±0.6°C ±1.1°F, ±0.05% of rdg ±0.8°C ±1.4°FDisplay resolution: 0.05°C 0.09°F*3

Measurement range - T

Measurable range: −100 to +400°C -148°F to +752°F, −200 to −100°C -328°F to -148°FMeasurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±0.5°C ±0.9°F, ±0.05% of rdg ±0.8°C ±1.4°FDisplay resolution: 0.01°C 0.02°F*3

Measurement range - N

Measurable range: 0 to 1300°C 32°F to 2372°FMeasurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±0.6°C ±1.1°FDisplay resolution: 0.05°C 0.09°F*3

Measurement range - C

Measurable range: 1500 to 2315°C 2732°F to 4199°F, 0 to 1500°C 32°F to 2732°FMeasurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±1.1°C ±2.0°F, ±0.05% of rdg ±0.8°C ±1.4°FDisplay resolution: 0.05°C 0.09°F*3

Measurement range - R

Measurable range: 300 to 1768°C 572°F to 3214.4°F, 0 to 300°C 32°F to 572°FMeasurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±0.8°C ±1.4°F, ±0.05% of rdg ±1.6°C ±2.9°FDisplay resolution: 0.05°C 0.09°F*3

Measurement range - S

Measurable range: 300 to 1768°C 572°F to 3214.4°F, 0 to 300°C 32°F to 572°FMeasurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±0.9°C ±1.6°F, ±0.05% of rdg ±1.6°C ±2.9°FDisplay resolution: 0.05°C 0.09°F*3

Measurement range - B

Measurable range: 400 to 600°C 752°F to 1112°F, 600 to 1820°C 1112°F to 3308°FMeasurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±1.7°C ±3.1°F, ±0.05% of rdg ±1.0°C ±1.8°FDisplay resolution: 0.05°C 0.09°F*3

Measurement range - Pt100

Measurable range: −200 to +660°C -328°F to +1220°FMeasurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.1% of rdg ±0.3°C ±0.5°FDisplay resolution: 0.02°C 0.04°F*4

Measurement range - JPt100

Measurable range: −200 to +510°C -328°F to +950°FMeasurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.1% of rdg ±0.3°C ±0.5°FDisplay resolution: 0.02°C 0.04°F*4

Reference contact compensation

Switchable between internal and external (individual unit setting, 0°C 32°F for external)

Reference contact compensation accuracy

±0.7°C ±1.3°F (23°C ±2°C 73.4°F ±3.6°F, at input terminal temperature equilibrium)

Maximum input voltage

±60 V (rated)

Input impedance - Thermocouple, Voltage: 5 V range or lower

10 MΩ or more

Input impedance - Voltage: 50 V, 10 V range

Approx. 1 MΩ

Withstand voltage - Measurement input terminal - System bus

1500 VAC (50/60 Hz) for 1 minute

Withstand voltage - Mutually between measurement input terminals

120 Vp-p AC/DC

Input signal source resistance - Voltage/thermocouple

2 kΩ or less

Input signal source resistance - Resistance thermometer

10 Ω or less per line (three lines must be equal)

Thermocouple burnout

Wire breakage detected in the thermocouple range by applying 1 μA of current

Burnout detection frequency

Detected in the measurement cycle

Buffer memory

200k data

Warm up time

30 minutes or more*5

Rating

1.2 W or less

Environmental resistance - Ambient temperature

0 to +40°C 32°F to 104°F

Environmental resistance - Relative humidity

10 to 85% RH (no condensation)

Weight

Approx. 230 g 8.12 oz (including approx. 100 g 3.53 oz for the OP-51582 (NR-TH08 terminal block))

Customer Reviews

0 reviews
0
0
0
0
0

There are no reviews yet.

Be the first to review “Keyence NR-TH08”

Your email address will not be published. Required fields are marked *